Advanced functions
70
LFXG-H Installation and Operation Guide
• Temperature probe
• Flow meter
• Second transmitter
While in this mode, after entering a number of counts, it may be useful to
look at the
Process Chain
tab to view the variables affected by the auxiliary
input counts value.
Procedure 4.7: To perform an auxiliary input test
Relay Test
This mode manually toggles the relay On or Off to test the contacts. This is
useful for verifying whether alarm annunciators are functioning.
Procedure 4.8: To perform a relay test
Temperature Test
This mode manually forces the sensor’s temperature probe output to a
specified value. This is useful for verifying the scintillator sensor
temperature compensation.
1.
Select
Diagnostics | Test | Auxiliary Input Test
.
2.
Click
Enter
.
3.
Remove the gauge from control. Enter the auxiliary counts.
4.
Click
OK
.
The transmitter functions in this mode until it times out (1 hour), or
you click
Exit
and
OK
.
1.
Select
Diagnostics | Test | Relay Test
.
2.
Select
Energize relay
or
De-energize relay
.
3.
The transmitter functions in this mode until it times out (1 hour), or you
click
Exit
.
Summary of Contents for FiberFlex LFXG-H
Page 2: ......
Page 4: ...ii LFXG H Installation and Operation Guide NOTES...
Page 6: ...iv LFXG H Installation and Operation Guide...
Page 14: ...Preface xii LFXG H Installation and Operation Guide...
Page 34: ...Installation 32 LFXG H Installation and Operation Guide FiberFlex mounting...
Page 91: ...Diagnostics and repair LFXG H Installation and Operation Guide 89 Source Functions...
Page 100: ...98 LSGH Installation and Operation Guide NOTES...
Page 101: ......