Vectra Genisys® Therapy System
46
OPErATiON
Give Patient Interrupt Switch to patient
and explain that pressing the Red button
once pauses the therapy session .
gENErAL ELECTrOThErAPY wAvEFOrM SET uP (CONTiNuEd)
Optional Patient Interrupt Switch
Set intensity by rotating the Intensity Control
Knob to the prescribed level .
Set Waveform Intensity
PrESSInG buTTOn OncE PauSES SESSIOn
InTEnSITy
DISPlayED
rOTaTE
InTEnSITy
knOb
Clockwise-
Increases Intensity
Counterclockwise-
Decreases Intensity
Intensity Knob Rotation
If Patient Interrupt Switch is depressed, the
treatment will be paused and a message
will appear on the System screen .
Press any button to clear the message .
NOTE:
If the Patient Interrupt Switch
is depressed a second time,
the message will clear from
the screen and the treatment
will remain paused .
Start Treatment
Press the Start button to begin therapy
session .
STarT
buTTOn