Vectra Genisys® Therapy System
128
iNSTALLATiON/rEMOvAL
Install all cables and accessories to the Front
Access Panel . Refer to
page 16
for Symbol
Definitions .
I
NSTALLiNg
s
EMg MOduLE (CONTiNuEd)
Install Cables and Accessories
Plug Power Cord into an approved outlet .
Turn the System On using the On/Off
Switch . The System will automatically
recognize the added Module and display a
configuration change message .
Read and carefully follow the instructions
on the Screen .
cOnnEcT all cablES
anD accESSOrIES
Apply Mains Power
Verify that the Module installed is the Module
displayed in the message BEFORE pressing
the START Button . If it is not, DO NOT press
the START Button . Turn the System OFF and
back ON . If the problem continues, call the
selling dealer or DJO, LLC Technical Support
immediately .
DO NOT USE THE SYSTEM until all necessary
repairs are made by a Technician certified
by DJO, LLC . If use is attempted before
repairs are made, the System may operate
unpredictably and has the potential of
causing injury to the patient or damage to
the System's internal components .