Vectra Genisys® Therapy System
82
OPErATiON
PATiENT dATA CArd- SET uP OF NEw CArd (CONTiNuEd)
The following information uses an IFC
Treatment as an example . Electrode
Placement procedures for all modalities are
performed in the same basic fashion .
Press the Electrode Placement button .
NOTE:
When Ultrasound is the modality,
only the Side button is available .
Access Electrode Placement
ElEcTrODE PlacEMEnT buTTOn
Electrode Placement Set Up
PaD
buTTOn
chan
buTTOn
SIDE
buTTOn
SIzE
buTTOn
Press the Up, Down, Left and Right Arrow
buttons to position the selected electrode
as close to the actual treatment location as
possible .
Press the Pad button to select the other
electrode . Repeat above procedure for
electrode positioning .
If applicable, press the Chan button to
select another channel and repeat above
procedures .
After positioning the electrodes, press the
Accept and Return Arrow button .
Electrode Placement
accEPT anD rETurn
arrOw buTTOn
ElEcTrODES
POSITIOnED
Press the Pad button to select the Red (+) or
Black (-) electrode .
Press the Chan button to select Channel 1
or Channel 2 .
Press the Side button to select Front, Back,
Left, or Right of the body graphic .
Press the Size button until desired electrode
size is displayed . If electrode desired is not
listed, select Other .