Vectra Genisys® Therapy System
96
OPErATiON
Give Patient Interrupt Switch to patient
and explain that pressing the Red button
once pauses the therapy session .
Set intensity by rotating the Intensity Control
Knob to the prescribed level .
Set Modality Intensity
PrESSInG buTTOn OncE PauSES SESSIOn
InTEnSITy
DISPlayED
rOTaTE
InTEnSITy
knOb
Clockwise-
Increases Intensity
Counterclockwise-
Decreases Intensity
Intensity Knob Rotation
If Patient Interrupt Switch is depressed, the
treatment will be paused and a message
will appear on the System screen .
Press any button to clear the message .
NOTE:
If the Patient Interrupt Switch
is depressed a second time,
the message will clear from
the screen and the treatment
will remain paused .
Make certain the Patient Interrupt Switch,
for the channel(s) being used, is connected
to the Therapy System . Refer to
page 16
for
Symbol Definitions .
NOTE:
When reinstalling the Front Access
Panel, make certain the Lanyard
does not become kinked .
Optional Patient Interrupt Switch
lanyarD
PaTIEnT
InTErruPT
SwITch
CLiNiCAL rESOurCES LiBrArY- CLiNiCAL PrOTOCOLS™ (CONTiNuEd)