Vectra Genisys® Therapy System
SET uP
30
Press the Display Unit Version
Information button to show
the system software versions
installed .
Press the Back button to return
the Operator Utilities screen .
PrESS DISPlay unIT
vErSIOn buTTOn TO vIEw
SOfTwarE vErSIOnS
back
buTTOn
ThErAPY SYSTEM SET uP (CONTiNuEd)
Display Unit Version Information
The Pad Contact Quality feature indicates to the user the contact
quality of the electrodes on the patient . This function, if On,
displays a bar graph at the bottom of Treatment Review screen for
the following waveforms only:
The Therapy System Pad Contact Quality default is ON .
PaD cOnTacT QualITy
buTTOn
Pad Contact Quality
Interferential:
•
Dual Channel Graph
IFC Premod (2p):
•
Single Channel Graph
Russian:
•
Single Channel Graph
To turn On or Off, press Pad
Contact Quality button until On
or Off as desired is displayed .
Single Channel Waveforms will
display a single bar graph . Dual
Channel waveforms will display a
double bar graph .
Contact quality is measured by
the amount of the graph filled
with black .
An ideal contact quality is 75% or
more of the graph filled .
SInGlE channEl GraPh
Dual channEl GraPh
GOOD cOnTacT
QualITy
channEl 1:
GOOD cOnTacT QualITy
channEl 2:
nO cOnTacT QualITy