Vectra Genisys® Therapy System
123
iNSTALLATiON/rEMOvAL
Remove the Therapy System from the
Therapy System Cart, if equipped . Refer
to
pages 131 through 133
for proper
instructions .
Place Therapy System face down on a clean
working surface protected with a soft, clean
fabric to prevent damage to the lens .
rEMOvAL- ChANNEL 3/4 ELECTrOThErAPY, NiMh BATTErY, ANd LASEr MOduLE
Disconnect Mains Power
Disconnect the system from the power
source (outlet or remove battery module
if installed) before attempting any
maintenance, installation, removal, or
replacement procedures to prevent electrical
shock and possible damage to system .
Disconnect the Mains Power Cord from
the power supply . Remove the Rear Access
Panel and disconnect the Mains Power
Cord from the Therapy System .
rEar
accESS
PanEl
MaInS
POwEr
cOrD
Remove Therapy System from Cart
NOTE:
It is not necessary to remove
the sEMG Module from the Channel
3/4 Electrotherapy Module, if installed .
Remove the Front Access Cover and
disconnect the Lead Wires and Accessories
from the Therapy System and Channel 3/4
Electrotherapy Module .
Remove Lead Wires and Accessories
rEMOvE lEaD
wIrES anD
accESSOrIES
NOTE:
Keep Lead Wires and accessories
for later re-installation to the
Therapy System .