9. Acquiring Dental CT Images
PCT-90LH User Manual
113
EN
G
LISH
9.5
X-ray Exposure
If an emergency occurs during image acquisition, release the
Exposure Switch
to cease X-ray emission.
The operator shall observe the X-ray safety regulations applicable
to their area at all times during the operation of this equipment.
During image acquisition (X-ray exposure), don’t use the Chinrest
UP/DOWN button or switch except for emergency situations. (The
movement of the Rotating Unit is necessary only for the patient
positioning in normal situation.) Doing so may lead the patient to be
unsafe.
The operator must keep vocal/visual contact with the patient at all
times during image acquisition process.
Do not operate the PC during exposure. Doing so may cause the
system to malfunction.
Let the patient close the eyes during operation.
To acquire optimized images, let the patient hold his/her breath
and not to swallow. Also don’t let the patient move until the
Rotating Unit moves up.
1.
Get out of the X-ray room and close the door.
The operator must keep vocal/visual contact with the patient at all
times during image acquisition.
2.
Press and hold down the
Exposure Switch
until image acquisition is completed.
The image appears on the screen.
Summary of Contents for Premium PAX-i3D
Page 1: ......
Page 2: ...PCT 90LH User Manual 3...
Page 27: ...4 Imaging System Overview PCT 90LH User Manual 21 ENGLISH 4 4 Imaging System Configuration...
Page 29: ...4 Imaging System Overview PCT 90LH User Manual 23 ENGLISH 4 5 Equipment Overview...
Page 44: ...4 Imaging System Overview 38 PaX i3D Green Premium User Manual Left blank intentionally...
Page 52: ...5 Imaging Software Overview 46 PaX i3D Green Premium User Manual Left blank intentionally...
Page 58: ...6 Getting Started 52 PaX i3D Green Premium User Manual Left blank intentionally...
Page 122: ...9 Acquiring Dental CT Images 116 PaX i3D Green Premium User Manual Left blank intentionally...
Page 146: ...11 Acquiring 3D PHOTOs Optional 140 PaX i3D Green Premium User Manual Left blank intentionally...
Page 148: ...12 Troubleshooting 142 PaX i3D Green Premium User Manual Left blank intentionally...
Page 152: ...13 Cleaning and Maintenance 146 PaX i3D Green Premium User Manual Left blank intentionally...
Page 154: ...14 Disposing of the Equipment 148 PaX i3D Green Premium User Manual Left blank intentionally...
Page 166: ...15 Technical Specifications 160 PaX i3D Green Premium User Manual Left blank intentionally...
Page 189: ...16 Appendices PCT 90LH User Manual 183 ENGLISH...
Page 204: ......