16. Appendices
168
PaX-i3D Green Premium™ User Manual
16.2.2 Leakage Dose
X-ray Leakage Dose Test is performed to protect patients against excessive and
unnecessary radiation that is not purposed and this document evaluates leakage dose
amount based on following standard defined by IEC regulation and has been
performed by covering each collimator region in use.
Standard
National
Deviation
Terminology
Permissive Range
International
Standard
IEC 60601-1-3
Leakage
limits leakage at 1M from the source to
100 mR in 1hr
Test Condition
Test mode
kVp
mA
sec
PANO / Adult
90
9.6
13.5
Dental CT (21 x 19)
110
6.5
18.0
ENT CT (21 x 19)
110
6.5
18.0
i-CEPH
102
4.5
18.0
Measurement Overview
Summary of Contents for Premium PAX-i3D
Page 1: ......
Page 2: ...PCT 90LH User Manual 3...
Page 27: ...4 Imaging System Overview PCT 90LH User Manual 21 ENGLISH 4 4 Imaging System Configuration...
Page 29: ...4 Imaging System Overview PCT 90LH User Manual 23 ENGLISH 4 5 Equipment Overview...
Page 44: ...4 Imaging System Overview 38 PaX i3D Green Premium User Manual Left blank intentionally...
Page 52: ...5 Imaging Software Overview 46 PaX i3D Green Premium User Manual Left blank intentionally...
Page 58: ...6 Getting Started 52 PaX i3D Green Premium User Manual Left blank intentionally...
Page 122: ...9 Acquiring Dental CT Images 116 PaX i3D Green Premium User Manual Left blank intentionally...
Page 146: ...11 Acquiring 3D PHOTOs Optional 140 PaX i3D Green Premium User Manual Left blank intentionally...
Page 148: ...12 Troubleshooting 142 PaX i3D Green Premium User Manual Left blank intentionally...
Page 152: ...13 Cleaning and Maintenance 146 PaX i3D Green Premium User Manual Left blank intentionally...
Page 154: ...14 Disposing of the Equipment 148 PaX i3D Green Premium User Manual Left blank intentionally...
Page 166: ...15 Technical Specifications 160 PaX i3D Green Premium User Manual Left blank intentionally...
Page 189: ...16 Appendices PCT 90LH User Manual 183 ENGLISH...
Page 204: ......