16. Appendices
178
PaX-i3D Green Premium™ User Manual
a
Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in
the location where the
PCT-90LH
is used exceeds the applicable RF compliance
level above, the
PCT-90LH
should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as
re-orienting or relocating the
PCT-90LH
.
b
Over the frequency range 150kHz to 80MHz, field strengths should be less than
[V
1
] V/m.
Recommended separation distances between portable and mobile RF
communications equipment and the PCT-90LH
This is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the
PCT-90LH
can help
Prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the
PCT-
90LH
as recommended below, according to the maximum output power of the
communications equipment.
Separation distance according to frequency of transmitter [m]
IEC 60601-1-2
Frequency of
Transmitter
Equation
Rated maximum
output power of
transmitter
[W]
150kHz to 80MHz
V
1
=3Vrms
Separation
Distance (meters)
80MHz to 800MHz
E
1
=3V/m
Separation
Distance (meters)
800MHz to 2.5GHz
E
1
=3V/m
Separation
Distance (meters)
0.01
0.116
0.1166
0.2333
0.1
0.368
0.3687
0.7378
1
1.166
1.1660
2.3333
10
3.687
3.6872
7.3785
100
11.660
11.6600
23.333
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters(m) can be estimated using the
equation applicable to the frequency of the transmitter, where p is the maximum
output power rating of the transmitter in watts(W) according to the transmitter
manufacturer.
Summary of Contents for Premium PAX-i3D
Page 1: ......
Page 2: ...PCT 90LH User Manual 3...
Page 27: ...4 Imaging System Overview PCT 90LH User Manual 21 ENGLISH 4 4 Imaging System Configuration...
Page 29: ...4 Imaging System Overview PCT 90LH User Manual 23 ENGLISH 4 5 Equipment Overview...
Page 44: ...4 Imaging System Overview 38 PaX i3D Green Premium User Manual Left blank intentionally...
Page 52: ...5 Imaging Software Overview 46 PaX i3D Green Premium User Manual Left blank intentionally...
Page 58: ...6 Getting Started 52 PaX i3D Green Premium User Manual Left blank intentionally...
Page 122: ...9 Acquiring Dental CT Images 116 PaX i3D Green Premium User Manual Left blank intentionally...
Page 146: ...11 Acquiring 3D PHOTOs Optional 140 PaX i3D Green Premium User Manual Left blank intentionally...
Page 148: ...12 Troubleshooting 142 PaX i3D Green Premium User Manual Left blank intentionally...
Page 152: ...13 Cleaning and Maintenance 146 PaX i3D Green Premium User Manual Left blank intentionally...
Page 154: ...14 Disposing of the Equipment 148 PaX i3D Green Premium User Manual Left blank intentionally...
Page 166: ...15 Technical Specifications 160 PaX i3D Green Premium User Manual Left blank intentionally...
Page 189: ...16 Appendices PCT 90LH User Manual 183 ENGLISH...
Page 204: ......