PaX-i (PCH-2500) User Manual
95
7. Acquiring CEPH Images
F. Place the ear rods along the patient’s
ear canals. Make sure that the ear rods
are comfortably, yet firmly in place.
G. Gently tilt the patient’s head back until
his/her Frankfurt plane is perpendicular to
the floor, as shown below.
H. Click the
Ready
button after the patient has been properly positioned. No
X-Ray will be emitted at this point. The CEPH sensor and the 2
nd
collimator will
enter positioning mode.
I. Proceed to section
7.4: Initiating X-Ray Exposure.
7.2.4
Waters View Mode
Follow the procedure below to prepare and position the patient for Waters View
mode image acquisition.
A. Widen the distance between the two
ear rods.
B. The nasal positioner should be flipped
to the side and up to prevent it from
obstructing the image acquisition.
C. Guide the patient to the CEPH unit.
D. Ask the patient to stand upright facing the sensor. Make sure that the
patient’s shoulders are level and that his/her neck is relaxed.
E. Adjust the height of the unit to suit the patient by pressing the column up/
down button or switch (optional).
CAUTION
After adjusting the column to suit the height of the patient,
fit the ear rods along the patient’s ear canals.
Frankfurt plane
Nasal positioner
Pi_u272_en_20180514.indd 95
2018-05-14 오전 11:03:29
Summary of Contents for PAX-I
Page 1: ...User Manual Full version Model PCH 2500 Version 2 72 English...
Page 2: ......
Page 38: ...This page is left intentionally blank...
Page 101: ...Troubleshooting 8...
Page 103: ...Cleaning and Maintenance 9 9 1 Cleaning 104 9 2 Maintenance 105...
Page 107: ...Disposing of the Unit 10...
Page 136: ......
Page 137: ...Postal code 18449 13 Samsung 1 ro 1 gil Hwaseong si Gyeonggi do Korea www vatech co kr...