PaX-i (PCH-2500) User Manual
91
7. Acquiring CEPH Images
CAUTION
After adjusting the height of the column to suit the patient,
fit the ear rods along the patient’s ear canals and adjust the
nasal positioner.
E. The ear rods should fit properly along
the patient’s ear canals. The patient’s
Frankfurt Plane should be parallel to the
floor.
F. Position the nasal positioner on the
patient’s Nasion point. The height of nasal
positioner can be adjusted.
G. Ask the patient to swallow saliva and remain still with his/her mouth closed
until X-Ray exposure is complete.
H. Click the
Ready
button after the patient has been properly positioned. No
X-Ray will be emitted at this point. The CEPH sensor and the 2
nd
collimator will
enter positioning mode.
I. Proceed to section
7.4: Initiating X-Ray Exposure.
Frankfurt plane
Pi_u272_en_20180514.indd 91
2018-05-14 오전 11:03:28