PaX-i (PCH-2500) User Manual
122
Appendices
2.
X-Ray Dose Data
The X-Ray dose data is extracted from the X-Ray Dose Test Report for the
PaX-i.
X-Ray Dose Test Report for the PaX-i maintains dosemetric evaluation of
VATECH dental diagnostic system meets all requirements specified in the IEC
Collateral Standard. To limit unnecessary exposure to the patient, operator or
other staff, the PaX-i is designed to comply with IEC 60601-1-3 Part 1 General
Requirements for Safety.
Test Condition
Model (Brand Name)
PCH-2500 (PaX-i)
Sensor type
PANO: Xmaru 1501CF
CEPH: Xmaru 2301CF
Tube type / Inverter type
EXG9
2.1
DAP Table
Testing Equipment
Instrument
Manufacturer
Model
S/N
Dose Meter
Piranha
255
CB2-08050121
Mode Tested: PANO HD Normal, 13.5 s
mA
4
5
8
[mGy]
[mGy
ㆍ
cm2]
[mGy]
[mGy
ㆍ
cm2]
[mGy]
[mGy
ㆍ
cm2]
60
5.01
33.5
7.79
52.0
10.26
68.4
70
7.12
47.5
10.41
69.5
13.72
91.6
80
9.06
60.5
13.26
88.5
17.48
116.7
90
11.12
74.2
16.28
108.7
21.45
143.1
kVp
Pi_u272_en_20180514.indd 122
2018-05-14 오전 11:03:32
Summary of Contents for PAX-I
Page 1: ...User Manual Full version Model PCH 2500 Version 2 72 English...
Page 2: ......
Page 38: ...This page is left intentionally blank...
Page 101: ...Troubleshooting 8...
Page 103: ...Cleaning and Maintenance 9 9 1 Cleaning 104 9 2 Maintenance 105...
Page 107: ...Disposing of the Unit 10...
Page 136: ......
Page 137: ...Postal code 18449 13 Samsung 1 ro 1 gil Hwaseong si Gyeonggi do Korea www vatech co kr...