PaX-i (PCH-2500) User Manual
79
6. Acquiring PANO Images
WARNING
Be careful not to project the laser beams directly into the
patient’s eyes as this could severely damage the patient’s
vision.
A.
Mid-sagittal plane laser beam:
Position the Mid-sagittal plane laser
beam at the center of the patient’s face to
prevent magnifications of the left or right
side in the final image.
B.
Frankfurt plane laser beam:
The
Frankfurt plane laser beam should go from
the top of the ear to the tip of the nose
when the patient’s head is tilted back 10 –
15 °.
You may adjust the Frankfurt plane laser
beam by manually adjusting the Frankfurt
plane laser beam lever.
C.
Canine laser beam:
Have the patient
smile to properly position the canine laser
beam on the center of the patient’s canine
tooth.
Completing Patient Positioning
A. After confirming patient positioning and beam alignment, adjust the temple
supports to fit snugly on both sides of the patient’s head using the temple
support wheel. The temple support wheel is located underneath the patient
support rest.
B. Click the
Ready
button on the PC or Touchscreen after the patient has been
properly positioned. At this point, X-Ray will not be emitted. The rotating unit
will move to position for imaging.
C. Proceed to section
6.4: Initiating X-Ray Exposure.
Canine laser beam
Mid-sagittal plane
laser beam
Adjusting lever for the
canine
Frankfurt plane laser
beam
Pi_u272_en_20180514.indd 79
2018-05-14 오전 11:03:26
Summary of Contents for PAX-I
Page 1: ...User Manual Full version Model PCH 2500 Version 2 72 English...
Page 2: ......
Page 38: ...This page is left intentionally blank...
Page 101: ...Troubleshooting 8...
Page 103: ...Cleaning and Maintenance 9 9 1 Cleaning 104 9 2 Maintenance 105...
Page 107: ...Disposing of the Unit 10...
Page 136: ......
Page 137: ...Postal code 18449 13 Samsung 1 ro 1 gil Hwaseong si Gyeonggi do Korea www vatech co kr...