PaX-i (PCH-2500) User Manual
127
Appendices
Guidance and manufacturer’s declaration - electromagnetic immunity
The model PaX-i(PCH-2500) is intended for use in the electromagnetic
environment specified below. The customer or the user of the model PaX-
i(PCH-2500) should assure that it is used in such an environment.
Emissions
test
Compliance
Compliance
level
Electromagnetic
environment -guidance
Electrostatic
d i s c h a r g e
(ESD)
IEC 61000-
4-2
± 6 k V
Contact ±8
kV air
± 6 k V
Contact ±8
kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30 %.
E l e c t r i c a l
f
a
s
t
t r a n s i e n t /
burst
IEC 61000-
4-4
± 2 k V f o r
power supply
lines ±1 kV
f o r i n p u t /
output lines
± 2 k V f o r
power supply
lines ±1 kV
f o r i n p u t /
output lines
M a i n s p o w e r q u a l i t y
should be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-
4-5
±1 kV line(s)
to line(s) ±2
kV line(s) to
earth
±1 kV line(s)
to line(s) ±2
kV line(s) to
earth
M a i n s p o w e r q u a l i t y
should be that of a typical
commercial or hospital
environment.
V o l t a g e
dips, short
interruptions
and voltage
v a r i a t i o n s
o n p o w e r
supply input
lines
IEC 61000-
4-11
< 5 % U
T
(> 95 % dip
i n U
T
) f o r
0.5cycle 40
% U
T
(60 %
dip in U
T
)
for 5 cycle,
6 c y c l e 7 0
% U
T
(30 %
dip in U
T
) for
25 cycle, 30
cycle <5 %
U
T
(< 95 %
dip in U
T
) for
5 s
< 5 % U
T
(> 95 % dip
i n U
T
) f o r
0.5cycle 40
% U
T
(60 %
dip in U
T
)
for 5 cycle,
6 c y c l e 7 0
% U
T
(30 %
dip in U
T
) for
25 cycle, 30
cycle <5 %
U
T
(< 95 %
dip in U
T
) for
5 s
M a i n s p o w e r q u a l i t y
s h o u l d b e t h a t o f a
typical commercial or
hospital environment.
If the user of the model
PaX-i(PCH-2500) image
i n t e n s i f i e r r e q u i r e s
continued operation during
power mains interruptions,
it is recommended that the
model PaX-i(PCH-2500)
b e p o w e r e d f r o m a n
u n i n t e r r u p t i b l e p o w e r
supply.
Pi_u272_en_20180514.indd 127
2018-05-14 오전 11:03:33
Summary of Contents for PAX-I
Page 1: ...User Manual Full version Model PCH 2500 Version 2 72 English...
Page 2: ......
Page 38: ...This page is left intentionally blank...
Page 101: ...Troubleshooting 8...
Page 103: ...Cleaning and Maintenance 9 9 1 Cleaning 104 9 2 Maintenance 105...
Page 107: ...Disposing of the Unit 10...
Page 136: ......
Page 137: ...Postal code 18449 13 Samsung 1 ro 1 gil Hwaseong si Gyeonggi do Korea www vatech co kr...