PaX-i (PCH-2500) User Manual
125
Appendices
2.3
X-Ray Scatter Dose
Test Equipment Information
Instrument
Manufacturer
Model
S / N
Dose Meter
Victoreen
660
101114 /
101377
PANO Mode
Test Condition
Mode Tested
PANO UHD
Distance from focal point [m]
1
Applied Tube Voltage Peak [kVp]
90
Applied Tube Current [mA]
10
Mode
Direction [°]
PANO UHD, 16.6 s [mR/hr]
1 m
(3.3 ft)
1.5 m
(4.9 ft)
2 m
(6.6 ft)
0
15.3
9.76
2.32
45
35.3
22.9
16.5
90
70.6
40.8
21.8
135
387
237
106
180
408
266
114
Mode
Direction [°]
PANO UHD, 16.6 s [mR/hr]
1 m
(3.3 ft)
1.5 m
(4.9 ft)
2 m
(6.6 ft)
225
166
63.9
38.8
270
29.5
13.4
7.74
315
19.6
10.1
6.33
Pi_u272_en_20180514.indd 125