PaX-i (PCH-2500) User Manual
89
7. Acquiring CEPH Images
7.2
Positioning the Patient
Follow the procedure below to prepare and position the patient for CEPH imaging.
Before Positioning the Patient
■
Ask the patient to remove all jewelry and metallic objects, such as earrings,
hairpins, spectacles, dentures, and orthodontic appliances. These items
may cause shadow images that could impede diagnosis.
■
It is strongly recommended that the patient wear a lead apron for protection
against any possible scatter radiation.
■
Adjust the column height to the patient’s height using the up/down button or
switch (optional).
UP
DOWN
Column up/down buttons on the
handle frame
Column up/down switch
(optional)
NOTE
Correct positioning is an important factor in capturing the
best possible image.
CAUTION
Ensure that the nasal positioner left unfolded, before
adjusting the ear rods in the proper direction.
WARNING
Ensure that the patient is clear of all moving parts while
adjusting the height of the equipment.
Pi_u272_en_20180514.indd 89
2018-05-14 오전 11:03:27