PaX-i (PCH-2500) User Manual
67
6. Acquiring PANO Images
H. A default value for the tube voltage (kVp)
and current (mA) will be displayed based
on the gender
on the gender and X-ray intensity. If necessary, further adjustments can be made
using the arrows to the right of each number.
NOTE
Adjustable resolution:
-
Tube voltage: ± 1 kVp
-
Tube current: ± 1 mA
I. Click the
Confirm
button for these
parameters to take effect.
NOTE
Please wait for a moment, while the rotating unit moves to
its initial scanning position.
When you click the
Confirm
button,
■
The
Ready
button will begin blinking to show that it has been activated.
This means that the equipment is ready for X-Ray exposure.
■
The rotating unit will move to its initial scanning position.
■
Three laser beams, Mid-sagittal plane, Frankfurt plane and Canine, will be
activated to facilitate patient positioning. These beams will disappear after
20 minutes or if the
Ready
button is clicked.
NOTE
The laser beam On / Off icons are located:
-
In the upper right side of the window:
-
-
On the Handle Frame:
■
The scan time and estimated DAP (Dose
Area Product) values will be shown
on the main display for the selected
exposure.
J. Guide the patient to the equipment. Position the patient within the equipment.
For further information about patient positioning, refer to section
6.3:
Positioning the Patient.
Pi_u272_en_20180514.indd 67
2018-05-14 오전 11:03:23
Summary of Contents for PAX-I
Page 1: ...User Manual Full version Model PCH 2500 Version 2 72 English...
Page 2: ......
Page 38: ...This page is left intentionally blank...
Page 101: ...Troubleshooting 8...
Page 103: ...Cleaning and Maintenance 9 9 1 Cleaning 104 9 2 Maintenance 105...
Page 107: ...Disposing of the Unit 10...
Page 136: ......
Page 137: ...Postal code 18449 13 Samsung 1 ro 1 gil Hwaseong si Gyeonggi do Korea www vatech co kr...