Model 979 Series Helium Mass Spectrometer Leak Detector
DRAFT 9/15/03
1-9
1.3 Services Required for Operation
1.3.1 Power
Depending upon the configuration you ordered, the mains power supply requirements for
the 979 Series leak detector are:
❑
100 VAC, 20 A, 50 Hz / 115 VAC, 20 A, 60 Hz, or
❑
230 VAC, 20 A, 50/60 Hz
CAUTION
Mains supply voltage fluctuations shall not exceed ±10% of the
nominal voltage.
Transient overvoltages shall not exceed values according to
Installation Category (Overvoltage Category) II of the
UL3101-1 Standard.
Fixed mains socket-outlet shall be connected to PROTECTIVE
EARTHING system.
For indoor use only, Pollution Degree II in accordance with
UL3101-1 Standard.
CAUTION
Vacuum Technologies strongly recommends the use of surge
protection to improve the immunity of the 979 Series leak
detectors against unidirectional transients caused by the
following phenomena:
❑
Switching phenomena in the power network
(e.g., switching of capacitor banks)
❑
Faults in the power network
❑
Indirect lightning strokes
WARNING
Use only the power cord that was provided with your leak
detector. The use of extension cords is not recommended and
could result in damage to the equipment and loss of warranty.