Model 979 Series Helium Mass Spectrometer Leak Detector
A-2
DRAFT 9/15/03
Figure A-2
Optically Isolated Output Circuit Sketch
Table A-1
Summary Table of Optically Isolated Outputs
Pin
Output Name
Momentary
or Level
Description
2
PWR
NA
Customer-su5 VDC to +24 VDC for all P101 outputs.
3
REJ 1
Level
Active when Leak Rate is greater than set point #1.
4
REJ 2
Level
Active when Leak Rate is greater than set point #2.
5
REJ 3
Level
Active when Leak Rate greater than set point #3.
6
REJ 4
Level
Active when Leak Rate greater than set point #4.
7
START OUT
Level
Active High when 979 in Rough Mode.
8
VENT OUT
Level
Active High when 979 in Vent Mode.
9
TEST OUT
Level
Active High when 979 in Test Mode.
10
BUSY OUT
Level
Active High when 979 in Busy Mode.
11
WAKE UP OUT
Level
Active High when 979 in Wake Up Mode.
12
NOT READY OUT Level
Active High when 979 in Not Ready Mode.
13
SPARE OUT
Level
Spare Output.
Customer
979 Series
5-24 VDC
I/O-2
I/O-3
I/O-4
I/O-12
I/O-13
PLC
+
Summary of Contents for 979 Series