Model 979 Series Helium Mass Spectrometer Leak Detector
B-10
DRAFT 9/15/03
EXTERNAL
Uses an external Calibrated Leak (in the test port). For Autocalibrate operations.
FPEAK
Adjusts the variable focus voltage to cause a maximum response to helium.
IDLE
Turns off the turbo and spectube and waits for a RUN command.
INCREMENT
Adds 1 to the Manual Ranging leak rate exponent (making it less negative).
Success is indicated by the normal ok response. Does nothing in auto-ranging
mode. Does nothing upon reaching the least sensitive range.
INTERNAL
Uses an internal Calibrated Leak. For calibrate operations.
KEEP
Causes the leak detector to close most valves and enter the HOLD state.
MANUAL
Sets to manual ranging mode. Success is indicated by the normal ok response.
NOSNIFF
Disables the internal sniffer mode.
ROUGH
Causes the leak detector to begin a roughing and testing sequence.
RUN
Starts the turbo and spectube after IDLE command.
SNIFF
Enables the internal sniffer mode.
SYTCATM
Sets the current system thermocouple reading to represent atmospheric
pressure. Success is indicated by the normal ok response.
SYTCZERO
Sets the current system thermocouple reading to represent a pressure that is
too low for a thermocouple to read. Success is indicated by the normal ok
response.
TPTCATM
Sets the current test port thermocouple reading to represent atmospheric
pressure. Success is indicated by the normal ok response.
TPTCZERO
Sets the current test port thermocouple reading to represent a pressure which is
too low for a thermocouple to read. Success is indicated by the normal ok
response.
TUNE
Adjusts the ion source voltage to cause a maximum response to helium.
VENT
Causes the leak detector to vent the test port to the atmosphere.
ZERO
Sets the current leak rate measurement to be 0.0 atm cc/sec in the most
sensitive range.
Table B-4
Leak Detection Actions (Continued)
Command
Action