Model 979 Series Helium Mass Spectrometer Leak Detector
DRAFT 9/15/03
xv
Solvents
CAUTION
The mechanical components of leak detectors are typically
cleaned with alcohol, methanol, or other solvents.
When heated, sprayed, or exposed to high-temperature
equipment, these solvents become flammable and explosive,
causing serious injury or death. Do not use these solvents near
a high-temperature source. Ventilate the working area with a
blower and work in a large, well-ventilated room.
Alcohol, methanol, or other solvents are irritants, narcotics,
depressants and/or carcinogens. Their inhalation and/or
ingestion may produce serious side effects. Prolonged or
continued contact with the skin will result in absorption
through the skin and moderate toxicity. Always ensure that
cleaning operations are carried out in large, well-ventilated
rooms, and wear eyeshields, gloves, and protective clothing.
CAUTION
Do not clean any aluminum parts with Alconox
®
. Alconox is
not compatible with aluminum and will cause damage.
NOTE
During reassembly, always use Loctite
®
PST
®
(teflon-impregnated pipe thread compound) on pipe threads.