Model 979 Series Helium Mass Spectrometer Leak Detector
DRAFT 9/15/03
xxi
Spectrometer Tube
CAUTION
The spectrometer tube operates at a very high vacuum
produced by the high vacuum pump. Service of the
spectrometer tube requires that this vacuum be vented to the
atmosphere.
CAUTION
Do not use grease or other substance on O-rings that will come
in contact with the spectrometer tube.
CAUTION
If the spectrometer tube magnet comes in contact with a
magnetic surface, the magnet may lose its gauss, causing the
spectrometer tube to lose sensitivity.
WARNING
Store the Ion Source in a cool, dry area in a tightly sealed
container. Wash hands thoroughly after handling the Ion Source
and especially before smoking or eating.
Gross Leak Option
If the Gross Leak option is installed, heed the following:
CAUTION
Do not make any alterations to the Gross Leak. Do not touch
the knurled nut on the Gross Leak.
The Gross Leak is calibrated at the factory and if any alteration
is made, the Gross Leak must be returned to the factory for
recalibration. For returns, contact Vacuum Technologies
Customer Service at 1-800-8VARIAN.
Pumps
WARNING
To avoid injury, use proper lifting techniques when moving
pumps. Your system may have pumps that require two people
to move them safely.