4-7
Model 979 Series Helium Mass Spectrometer Leak Detector
DRAFT 9/15/03
3. The spectrometer tube is located in the center front of the 979 just behind the front
panel (Figure 4-2).
Figure 4-2
Front View of the Spectrometer Tube Assembly
In addition to the three connectors on the top of the unit, indicated by
➀
in Figure 4-2,
and the ground cable
➁
, the spectrometer tube is attached by a KF-25 (ISO NW-25)
quick clamp
➂
and a wing nut located below the tube (not shown in this view).
4. Remove the three connectors from the top of the spectrometer tube (Figure 4-3). The
connector cables are labeled TC, Ion Source, and Preamplifier. If the cables are not
labeled, label them now for future reference.
Figure 4-3
Removing the Connectors
➀
➁
➂
➀