Model 979 Series Helium Mass Spectrometer Leak Detector
1-18
DRAFT 9/15/03
1.10 Specifications
The specifications for the 979 Series Leak Detector are provided in Table 1-1.
Table 1-1
979 Series Specifications
Specifications
979 Helium Mass Spectrometer Versions
System Sensitivity
Standard Sensitivity
High Sensitivity
High Vacuum Pump Type,
Air-cooled,
maintenance-free
Vacuum Technologies Turbomolecular Pump
Rough/Fore Pump
Wet (Oil-sealed)
Dry (TriScroll
)
Wet (Oil-sealed)
Dry (TriScroll
)
Rough Pump Type
Free Air Displacement
l/min, 60/50 Hz
—
SD-451
—
TS-620
—
SD-451
—
TS-620
—
410/342
—
500/420
—
410/342
—
500/420
Fore Pump Type
Free Air Displacement,
l/min, 60/50 Hz
DS-302
DS-302
TS-620
TS-320
DS-302
DS-302
TS-620
TS-320
285/237
285/237 500/420
250/210 285/237
285/237
520/420
250/210
Minimum Detectable Leak,
atm-cc/sec
5 x 10
–10
5 x 10
–11
Leak Units
atm-cc/sec, mbar-l/sec, torr-l/sec, Pa-m
3
/sec
Response Time
Less than 0.5 second, in accordance with AVS Standard 2.1.
Amplifier Drift
Less than 5% of full scale of the most sensitive range, in accordance with AVS
Standard 2.1.
Noise Level
Less than 5% of full scale, peak-to-peak, in accordance with AVS Standard 2.1.
Calibration
Fully automated tuning and calibration using the internal or external calibrated
leak. The fast calibrate feature allows the standard version to calibrate in less
than 20 seconds and the high sensitivity version to calibrate in less than
1 minute.
Zero Control
Selectable mode background elimination with unique AUTOZERO<0 feature.
Leak Indication
A 50-segment bar graph (linear or log) indicator displays leak rates
automatically; alphanumeric indication on LCD display. A leak can also trigger
an audible alarm frequency that varies in proportion to the size of the leak,
dedicated volume control and programmable audio threshold. Analog voltage is
proportional to the leak rate.