Models 947, 948 and 960 Leak Detectors Operations and Service Manual
A-7
[SHIFT]
ON
A SEQ
OFF
Turns on the Auto Sequencer function.
2
FIL SEL
1
Selects filament #1 in the Ion Source.
[SHIFT]
2
FIL SEL
1
Selects filament #2 in the Ion Source.
OUT V
SHOW
FIL #
Displays the active filament number.
[SHIFT]
OUT V
SHOW
FIL #
Displays the format of the analog leak rate output voltage.
[SHIFT]
RAW DTA
1
Displays a set of readings from the leak rate A/D conversion.
SET
REJ #
SHOW
Displays the active reject set point (1 through 4) and the value of
the reject set point.
[SHIFT]
SET
REJ #
SHOW
Sets the active reject set point. Enter a number from 1 through 4,
then press the function key.
SET
STD LK
SHOW
Displays the leak rate value to which the leak detector calibrates.
[SHIFT]
SET
STD LK
SHOW
Sets the leak rate value to which the leak detector calibrates. Enter
X.XE-X (the desired value), followed by the SHIFT key, then the
function key.
REJ3
SET
REJ 1
Sets the leak rate value for Reject Set Point #1. Enter X.XE-X
(the desired value), then the function key.
[SHIFT]
REJ3
SET
REJ 1
Sets the leak rate value for Reject Set Point #3. Enter X.XE-X
(the desired value), followed by the SHIFT key, then the
function key.
REJ 4
SET
REJ 2
Sets the leak rate value for Reject Set Point #2. Enter X.XE-X
(the desired value), followed by the SHIFT key, then the
function key.
[SHIFT]
REJ 4
SET
REJ 2
Sets the leak rate value for Reject Set Point #4. Enter X.XE-X
(the desired value), followed by the SHIFT key, then the
function key.