Models 947, 948 and 960 Leak Detectors Operations and Service Manual
4-37
4.2.10.3 Replacing the Test Port
To replace the test port:
1. Place the O-ring into the groove of the compression ring.
2. Place the compression ring and O-ring onto the top of the test port (Figure 4-15).
Figure 4-15
Placing the O-ring and Compression Ring onto the Test Port
3. Place the test port tightening nut onto the test port and turn it clockwise until it is just
tight enough to stay in place.
4. Place the test port plug into the test port tightening nut, making sure it is completely
seated.
5. Tighten the test port tightening nut completely.