Models 947, 948 and 960 Leak Detectors Operations and Service Manual
4-25
7.
Models 948/960 only:
Once the system has pumped down, clean and fill the Liquid
Nitrogen Cold Trap using the procedure described in “Liquid Nitrogen Cold Trap
(Models 948/960 only)” on page 4-10.
Follow all warnings and cautions in the procedure.
8. Calibrate and tune the 947/948/960 using the procedure described in “Re-calibrate the
Internal Calibrated Leak” on page 4-7.
9. Leak check the assembly before using the 947/948/960 for testing.