4343W/4336W-G5/2530W-G5 X-ray Detectors
30
www.vareximaging.com
R
E
FE
RE
NCE
M
ANUAL
11.7
Power-on and Power-off Sequence
To Power-on:
1
Insert the Battery into the 1 or 3-bay charger to remove from shut-down mode (only applies
to Batteries that are new).
2
Place Battery into X-ray detector making sure the Battery latches into place, see
Section 11.4.1
and
Section 11.4.2
. The X-ray detector will automatically
power-on
when battery is inserted.
Note
When a Battery is inserted into the X-ray detector, the LED Status Indicator
will turn orange as it boots. After booting, it connects to the configured
wireless router and is in Standby Mode, where the LED Status Indicator will
blink twice (2) per second. If the X-ray detector does not connect to the
configured wireless router, it will blink slowly.
To Power-off:
1
Remove the battery from the X-ray detector.
Note
Removal of the Battery does not automatically power-off the X-ray detector.
The X-ray detector will stay powered-on for approximately 3 minutes or
until discharged after battery removal.