4343W/4336W-G5/2530W-G5 X-ray Detectors
www.vareximaging.com
29
11.6
Using a Third-Party Inductive Battery Charger
Any third-party Inductive Battery Charger meeting the
Qi 1.2 standard at 15 watts
may be used
for battery charging outside of the bucky tray. Refer to
Figure 19
for the location of the inductive
charger receiver on the X-ray detector for integrating a third-party Inductive Battery Charger.
WARNING
Third-party Inductive Battery Chargers must not be used in the bucky
tray. Ignoring this warning may result in image artifacts or damage to
the X-ray detector.
Caution
Third-party Inductive Battery Chargers do not support image
acquisition while using the charger.
Note
To support image acquisition while using the charger, the Varex
supplied Inductive Battery Charger must be used, see
Section 11.5
.
Figure 19 Third-party Inductive Battery Charger 4343W/4336W-G5/2530W-G5 X-ray Detector