4343W/4336W-G5/2530W-G5 X-ray Detectors
Table 18
4336W-G5 and 2530W-G5 Radiated/Conducted Emissions, Harmonics, Voltage,
Fluctuations & Flicker
Emissions test
IEC 60601-1-2
test level
Compliance
Electromagnetic environment
RF conducted
emissions
EN55011/CISPR11
Group 1, Class A,
150 kHz – 30 MHz
For Group 1, in
frequency range
150KHz to 30 MHz
limits are not specified,
the test is unnecessary.
The detector uses RF energy for its
internal function. Nearby electronic
equipment may be affected.
RF radiated
emissions
EN55011/CISPR 11
Group 1, Class A,
30 MHz – 1 GHz
Group 1, Class A,
30 MHz – 1 GHz
The detector uses RF energy for its
internal function. Nearby electronic
equipment may be affected.
Harmonic
emissions
EN/IEC 61000-3-2
Class A
Class A
The detector is suitable for use in all
establishments other than domestic and those
directly connected to the low voltage power
supply network that supplies buildings used
for domestic purposes.
Voltage
fluctuations / flicker
emissions
IEC 61000-3-3
Complies
Complies
The detector is suitable for use in all
establishments other than domestic and those
directly connected to the low voltage power
supply network that supplies buildings used
for domestic purposes.
16.2
Electromagnetic Immunity
Table 19
4343W ESD, Transient/Burst, Surge, Voltage Variation, Magnetic Fields
Immunity test
IEC 60601-1-2
test level
Compliance
Electromagnetic environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
Contact Discharge: ± 2,
4, 8 kV
Air Discharge: ± 2, 4, 8,
15 kV
Contact Discharge: ±
2, 4, 8 kV
Air Discharge: ± 2, 4,
8, 15 kV
Floors should be wood, concrete, or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV AC Mains
±1 kV I/O Lines
N/A
Battery power
equipment not
connected to mains
Mains power quality should be that of a
typical professional healthcare
environment.
Surge
IEC 61000-4-5
±0.5 kV, ±1 kV
Line to Line
±0.5 kV, ±1 kV, ±2 kV
Line to Ground
N/A
Battery power
equipment not
connected to mains
Mains power quality should be that of a
typical professional healthcare
environment.