4343W/4336W-G5/2530W-G5 X-ray Detectors
11.13 Image Acquisition
Acquisition can be performed after Offset and gain calibration have been performed. This
section describes how to acquire images.
Before performing acquisition, settings should be checked to ensure that the desired corrections
are applied to acquired frames.
Table 14
lists the types of corrections that can be applied.
Table 14
Types of Corrections that can be applied during Acquisition
Type
Images transferred to Workstation
Raw
Exposed Frame and Post-Offset Frame
Corrected
A single fully corrected frame
Offset Corrected
A single offset corrected frame, without gain and defect correction
Gain Corrected
A single gain corrected frame, without defect correction
Note
When acquiring Corrected, Offset Corrected, and Gain Corrected images; if
preview is enabled and additional preview frame will be delivered to the
workstation prior to any additional frames.
1
To set the Correction type, click on
Settings
, the
Acquisition Settings Menu
will appear.
Figure 40 Radiographic Acquisition Settings