4343W/4336W-G5/2530W-G5 X-ray Detectors
www.vareximaging.com
v
Caution
It is important that the X-ray detector is not directly connected to the
installed network. Connection of the X-ray detector with the installed
computer network may disturb the IT environment.
Caution
Do not operate the X-ray detector in a location with the following
conditions:
•
Close to fluid or places where fluid is used
•
Close to heat sources, such as a heater
•
High temperature environment
•
High humidity environment
•
High condensation environment
•
Extreme cold environment
•
Dusty environment
•
Salty or sulfurous environment
•
Near a vibrating environment
•
Environment where there is insufficient air circulation to enable the X-ray
detector and power supply to dissipate heat
Ignoring this caution may result in personal injury or damage to
the product.
Caution
The X-ray detector is an applied part (patient contact device) and the
surface shall not exceed 42°C. Internal temperature sensor data is
provided in the diagnostic data attached to each image. These
temperature measurements are well correlated with the X-ray detector
external surface temperature. It is advisable to monitor this diagnostic
data as an additional safety precaution, see document
03236 FP2032 - VSP Software Interface. Also, see
Table 5
.
Note
The X-ray detector is intended to be installed, maintained, and used in a
professional healthcare facility.
Caution
Monitor the temperature of the X-ray detector and allow the X-ray
detector to cool down when the temperature of the X-ray detector gets
hot. The device is not designed to supply heat to a patient. Ignoring
this caution may result in personal injury or damage to the product.
Caution
An Image Quality Test or Customer Acceptance test should be
performed before the X-ray detector is used for the first time with
patients.
No
Caution
No part of the X-ray detector is intended to be attached to the patient
and/or contact the patient.
R
E
FE
RE
NCE
M
ANUAL