4343W/4336W-G5/2530W-G5 X-ray Detectors
26
www.vareximaging.com
R
E
FE
RE
NCE
M
ANUAL
11.5
Optional Varex Provided Inductive Battery Charger
The Battery may be charged using the Varex provided Inductive Battery Charger while the X-ray
detector is installed in the bucky tray.
Note
During image acquisition or calibration, inductive battery charging will
pause if Varex Inductive Battery Charger is used.
11.5.1
Installing the Varex Provided Inductive Battery Charger
1
Install the Varex Inductive Battery Charger and Detector Support to the bucky tray.
2
Self-tapping screws with washers, machine screws with washers into PEM nuts (installed on
the bucky tray), or 1mm (thickness) VHB Tape may be used to attach the Varex Inductive
Battery Charger and Detector Support into the bucky tray. Use
Figure 16
,
Figure 17
, and
Figure 18
for mounting locations.
3
Route the USB-C Cable from the Varex Inductive Battery Charger to the Power Supply in a
way that the opening and closing of the bucky tray does not cause damage to the cable. The
USB-C Cable can be replaced if damaged.
WARNING
Alignment between the Varex Inductive Battery Charger and the X-ray
detector should be exact. Poor alignment or foreign objects may cause
overheating.
Caution
The Varex Inductive Battery Charger and the X-ray detector charging
location should be flush when charging.
Caution
The Varex Inductive Battery Charger and Detector Support may not fit
all bucky trays. System Integrator should qualify bucky tray before
installing.
Caution
Screw heads must be sub-flush to avoid damage caused to the X-ray
detector when inserted into the bucky tray.