8.2.4 Measure centre
Here the TCAM2 determines the edge pixel next to the cross hair origin point. This is
done using the pointers in the adjustment bars that should, ideally, both be in the
middle. The edge point is next to the cross hair origin point when both pointers are in
the middle of the green area.
You access this operating mode using the following menu steps:
Main menu
Sub-menu 1
Imaging
Ö
Measure centre
Selecting this measuring method is recommended when carrying out the following
setup settings because any dust and dirt residue on the object to be measured, e.g.
gauge (calibrating mandrel or ball gauge), will not affect the measuring results.
•
Parallel correction
•
Absolute zero point