Output format
Counter
Axis 1
Ö
Axis 2
Corrections
Linear correction (1)
Ö
Parallel correction
Absolute zero point (1)
Options
Password
Date and time
Service
Factory settings
Ö
Delete tool references
Delete tools
5.1.2 Live picture
The object to be measured and the various adjustment aids depending on the operat-
ing mode (cross hairs, adjustment bars, test LED, measuring lines, analysis window,
measuring results) are displayed on the live picture operating mode.
5.1.2.1 Cross hairs
The cross hairs give an overview of the position of the contour.
In the "Fixed axis" operating mode the measurements always refer to these cross
hairs with the origin at 0.0 mm or inches. In the "Preset" and "Maximum" operating
modes these cross hairs remain so that the tool reference for the whole measuring
structure is retained. Additional red measuring lines are created on the object’s pro-
file depending on the measuring method determined or selected.
5.1.2.2 Adjustment bars
Adjustment bars are provided for both directions in the "Fixed axis" operating mode.
The pointer in the X adjustment bar gives the contour’s positive or negative deviation
from the measuring line in the X direction, the Z adjustment bar does the same in the
Z direction. The different colours on the bar show the distance of the contour from the
measuring line.
The different setting ranges have the following limits:
Colour Setting range
Grey ± (16 – 60) µm
Yellow ± (4 -15) µm
Green ± (0 -3) µm
5.1.2.3 Test LED
The test LED indicates whether the TCAM2 can correctly apply one of the imple-
mented measurement functions to the object displayed.