digiset
5
Urma AG, Switzerland www.urma.ch
page 13
4.2.2
Check illumination
After the working distance is checked, the homogeneity and brightness of the illumi-
nation must be set. For this, you
must not have
anything in the camera’s visual field.
To check the illumination the greyscales are checked in the camera’s horizontal and
vertical alignment on both edges of the image as well as the middle of the image and
the results shown on 6 lines.
The illumination reaches its optimum setting if these lines are as horizontal as possi-
ble and within the horizontal tolerance range marked by 2 purple lines.
You access the operating mode for setting the illumination using the following menu
steps:
Main menu
Sub-menu 1
Sub-menu 2
Setup
Ö
Imaging
Ö
Illumination
The TCAM2 displays the following image with 6 grey scales (illumination lines):
Procedure:
1. Select the illumination check with the "Illumination" operating mode. The
horizontal tolerance field appears with the illumination lines. "Illumination"
appears in the status window.
2. If not all 6 greyscale lines on the LCD are inside the tolerance field, you
have to adjust the light source. Remove the cover of the light source as
shown on the picture below.
Tolerance field