4.CALIBRATION
54
4-7. Secondary Calibration (Equivalent Calibration)
By key-inputting the weight value corresponding to the output value (mV/V) of the load cell,
calibration is performed so that the input of the entered output value results in the entered weight
value display.
Operation
1) Select setting mode 3.
2) Input the Balance Weight Value corresponding to the output value to be entered in 4).
(1-Balance Weight Value)
3) Select setting mode 9.
4) Input the Output value of the load cell. (2-Equivalent Calibration)
* If any error message appears, refer to "11.OVER SCALE & ERROR" on P.157.
* If “
” (Zero Calibration again) appears, redo Zero Calibration, and then perform
Equivalent Calibration again.
* “
” is displayed when calibrating by resolution that exceeds 1/10000.
→
→
→
F
CNG/ENT
CNG/ENT
3
NEAR Z
.
→
→
→
Balance Weight Value
(0 ~ 99999)
1
UPPER
CNG/ENT
CNG/ENT
→
→
→
F
CNG/ENT
CNG/ENT
9
FINAL
→
→
→
Output value of the load cell
(0.5000 ~ 3.2000)
2
LOWER
CNG/ENT
CNG/ENT