4.CALIBRATION
43
4-3. Secondary Calibration Procedure (Equivalent Calibration)
Calibration procedure performed by entering the rated output value (mV/V) and rated capacity
value of load cell without using real load.
Secondary calibration function is provided for provisional calibration when the F701-C develops
trouble or the calibration value is mistakenly changed.
The secondary calibration is only a provisional method. Calibration with actual load must be done
as soon as possible.
Release calibration inhibit LOCK by the switch OFF on the rear
panel.
Release Setting Value LOCK which inhibits the calibration.
The displayed unit is set.
Register the position of decimal point.
Register the maximum capacity of the scale.
If the registered value exceeds by 9 scale divisions, display shows
over scale, “
”.
Register the minimum unit (scale division) of the scale.
Set the weight value corresponding to the output value to be
entered in *1.
Register the rated output value (mV/V) of load cell.
If it is necessary, re-calibrate initial zero point.
Disable changing setting value related to calibration.
Turn the calibration LCOK ON to prohibit the calibration for
avoiding false operation.
Release Calibration LOCK
Decimal Place
Min. Scale Division
Capacity
Balance Weight Value
Equivalent Calibration
Zero Calibration
Calibration LOCK
LOCK Switch OFF
LOCK Switch ON
*1
Unit