4.CALIBRATION
41
4. CALIBRATION
4-1. Span Calibration
Calibration is performed for matching the F701-C to a load cell. For example, it is work to adjust so
that the F701-C accurately displays 100.00kg when an actual load (or weight) of 100kg is applied to
the load cell (balance section) of the weighing apparatus to which the F701-C is connected. This
operation is called Span Calibration.
Connect F701-C to the load cell.....
?
?
?
The display is incorrect.....
100kg
100kg
After calibration.....
The F701-C and load cell
function as a weighing device.