9.EXTERNAL INPUT/OUTPUT SIGNALS (CONTROL CONNECTOR)
122
9-5-10. Complete
◆
In simple comparison mode
Select the complete signal output timing in complete signal output mode (Weighing Function 2 in
setting mode 2).
The output ON time depends on the Complete Output Time in setting mode 1.
* For the method of setting, refer to "6-8.Complete Signal Output Mode / Complete Output Time /
Judging Time / Comparison Inhibit Time / Output Selection 2" on P.93.
◆
In sequence mode
- With Over/Under judgment
Conditions: The setting of Final and Over/Under comparison is
other than “comparison OFF”.
The number of Judging Times is other than 0.
Select the complete signal output timing in complete signal output mode (Weighing
Function 2 in setting mode 2).
The output ON time depends on the Complete Output Time in setting mode 1.
- Without Over/Under judgment
Conditions: The setting of Final and Over/Under comparison is
“comparison OFF”.
The number of Judging Times is 0.
The complete output turns ON in defiance of the setting of complete signal output mode
(Weighing Function 2 in setting mode 2) on the OFF edge (ON
→
OFF) of the set point 3
signal.
The output ON time depends on the Complete Output Time in setting mode 1.
* The complete output signal is used for go output by setting (output selection 2 under Output
Selection in setting mode 5).
(Either go or complete is selectable.)
* For the method of setting, refer to "6-8.Complete Signal Output Mode / Complete Output Time /
Judging Time / Comparison Inhibit Time / Output Selection 2" on P.93.
- Unless the weight value once drops down to 25% or less of the Final after the
complete output, the complete output will not turn ON after the next weighing.
- If the Final is set at 0, the complete output may turn ON at power-on.