2.CONNECTION
8
2. CONNECTION
2-1. Load cell Connection
The voltage application of F701-C is 10V, and the maximum current is 120mA, to which up to four
350
Ω
load cells can be connected in parallel.
Load cell terminal block pin assignments
Pin No.
Signal (6-wire)
Signal (4-wire)
1
+SIG
+SIG
2
-SIG
-SIG
3
+EXC
+EXC
(Connect 3 to 4)
4
+S
5
-EXC
-EXC
(Connect 5 to 6)
6
-S
7
SHIELD
SHIELD
Load cell terminal block