Appendix
79/92
data“ (JH-Paper).
Otherwise w
θ
and w
k
in equation (3) can be
rated manually with parameters w
θ
and w
k
.
w
θ
FLOAT
Rating for the retention data (only considered
if IN_WEIGHTFLAG = FALSE.)
wk FLOAT
Rating of the conductivity data (only
considered if IN_WEIGHTFLAG = FALSE.)
The parameter collections “Control parameters for Levenberg-
Marquardt algorithm“, “Control parameters for SCEUA algorithm“ and
“LM-Floatpar” should NOT show up. These are special parameters for
setting the optimization algorithm.
11.1.2 Output
RMSE
θ
FLOAT
Root mean squared error of the retention
function adaptation
RMSE
K
FLOAT
Root mean squared error of the conductivity
function adaption
,
with
and , resp.
and being the measured resp. predicted
values for the retention and conductivity function. n
p
is the number of
data points.