Set-up the HYPROP
39/92
Please note the following requirements for the operation of the scale
(also see 6.1 Starting conditions p 48):
1. The scale should be placed on a vibration-free work table.
2. The work table should only be used for the HYPROP
measurement.
3. The scale must be leveled out. Most scales have a bubble-level.
5.2 Software tensioVIEW
®
5.2.1 Menu
tensioVIEW
®
has simple, mostly self-explaining menus for read-out
and configuration of tensioLINK devices.
After starting tensioVIEW
®
the display is more or less blank, most
functions are not activated.
5.2.1.1 Find devices
If one or more sensors are connected via the USB-converter
they can be found by pressing the “magnifying glass” button.
tensioVIEW
®
offers two options for searching:
5.2.1.2 Single device mode
tensioVIEW
®
expects that only one device is connected which
will be found very quickly. This mode is not functional if more
than one device is connected!
5.2.1.3 Multiplexed devices mode
tensioVIEW
®
is able to run up to 20 HYPROP sensor units
connected to the bus within 8 seconds, but only if each device
is already personalized with an individual bus identification address.
If two or more devices have an identical address, none of them will
be found.
All devices found will be displayed in the left section in a directory
tree. Same types of devices will be grouped in one directory.