42
SECTION 6
OPERATION
6.5
Variable Speed Operation
When using the variable speed function to change the pump rotation rate, use a rotation rate
that does not cause resonance with other devices installed at the site.
The variable rotation speed function is only available by remote control or RS-485 operation.
Rotation speed cannot be varied by local control.
6.5.1 Outline
(1) The rotational speed settings function sets the rotational speed by selecting between the
NORMAL speed mode or LOW SPEED mode.
(2) Select the NORMAL mode or LOW SPEED mode by REMOTE operation using the "LOW
SPEED" signal input (Refer to Table 6-6).
(3) The NORMAL mode or LOW SPEED mode selection can be made before or after start-up.
(4) Set the low speed value between 25 % and 100 % of the rated speed in 0.1 % increments,
using the RS-485 communication.
(5) The low speed value can be set while the pump is rotating in the LOW SPEED mode. The
pump then accelerates or coasts to the new set value and maintains the set speed.
(6) The time required for the speed to change is the same as the time for normal acceleration
or coasting.
For example, if the low speed value is set to 80 % and the LOW SPEED mode is selected
during normal rotation, the time for the speed to drop to 80 % is approximately one-fifth the
time required to stop from rated speed.
CAUTION
NOTICE
Summary of Contents for UTM300B
Page 8: ...vii Location of label...
Page 11: ...x This page is intentionally left blank...
Page 17: ...xvi Table of contents This page is intentionally left blank...
Page 51: ...34 SECTION 5 INSTALLATION This page is intentionally left blank...
Page 70: ...GAS PURGE 7...
Page 77: ...60 SECTION 8 TURBO MOLECULAR PUMP RECONDITION This page is intentionally left blank...
Page 111: ...A 26 Appendix A COMMUNICATIONS This page is intentionally left blank...