A-7
A4
TABLE OF COMMANDS
Type
Command /
answer
Name
Command
character
string
Sub-command character string
History 1
Command
Read alarm history
GA
aa
※
9
Answer
Send alarm history
GB
xxx...xxx
※
10
No history data
GV
aa
※
9
History 2
Command
Read alarm history
GJ Aa
※
9
Answer
Send alarm history
GK xxx…xxx
※
19
No history data
GV aa
※
9
Settings
Command
Read settings
SR aa
※
11
Write settings
SW aabbbb
※
12
Answer
Send settings value
SA aabbbb
※
12
Invalid setting number
SV aa
※
11
Command
Read user memo
SU
None
Write user memo
SX xxx…xxx
※
13
Answer
Send user memo
SF xxx…xxx
※
13
Command
Default setting
SG
None
※
17
Answer
Execute
SH
None
RS-485 settings
Command
Read setting value
DR aa
※
14
Write setting value
DW aabbbb
※
15
Answer
Transmit setting value
DA aabbbb
※
15
Invalid setting value
number
DV aa
※
14
Command
Default setting
DD
None
※
18
Answer
Execute
DB
None
Shared answer
Answer
Invalid command
AN
None
※
1
aa : Failure alarm codes (hexadecimal) corresponding to the protection system.
Refer to Table 9-6 "Table of Alarms" and Table 9-7 "Table of Warnings".
※
2
aa : Parameter number (decimal). Refer to Table A-4 "Table of Parameters".
※
3
aa : Parameter number (decimal). Refer to Table A-4 "Table of Parameters".
bbbb : Parameter value (decimal). Refer to Table A-4 "Table of Parameters".
※
4
aa : Alarm list number (decimal)
※
5
aa : Alarm list number (decimal)
bb : Alarm code (decimal) stored in the alarm list with the requested number.
Refer to Table 9-6 "Table of Alarms" and Table 9-7 "Table of Warnings".
※
6
aa : Timer number (decimal). Refer to Table A-5 "Table of Timer".
※
7
06 (fixed) : Timer number (decimal). Refer to Table A-5 "Table of Timer".
aaaaa : Set value (decimal). Refer to Table A-5 "Table of Timer".
※
8
aa : Timer number (decimal). Refer to Table A-5 "Table of Timer".
bbbbb : Timer value (decimal). Refer to Table A-5 "Table of Timer".
cccccccccc : Time when the timer updated. (YYMMDDHHMM format. Stored as Greenwich
Mean
Time)
dddddddddd : Time when the timer reset. (YYMMDDHHMM format. Stored as Greenwich
Mean
Time)
※
9
aa : History number
※
10
xxx...xxx : Refer to Table A-6 "Alarm History 1 Data Format".
※
11
aa : Settings number (decimal). Refer to Table A-7 "Table of Settings".
Summary of Contents for UTM300B
Page 8: ...vii Location of label...
Page 11: ...x This page is intentionally left blank...
Page 17: ...xvi Table of contents This page is intentionally left blank...
Page 51: ...34 SECTION 5 INSTALLATION This page is intentionally left blank...
Page 70: ...GAS PURGE 7...
Page 77: ...60 SECTION 8 TURBO MOLECULAR PUMP RECONDITION This page is intentionally left blank...
Page 111: ...A 26 Appendix A COMMUNICATIONS This page is intentionally left blank...