RF POWER SUPPLY and MATCHINGBOX
5. Operation
67
RF-ON/OFF
EXT.CONT(1-2)
PRG STROBE
EXT.CONT(33-31)
ACK
EXT.CONT(42-43)
PRG SELECT
EXT.CONT
(27/28/29/30-31)
MOTO STROBE
EXT.CONT(32-31)
TUNE/MATCH
Position Monitor
EXT.CONT(44-45/46-47)
ON
OFF
CLOSE
OPEN
0V
10V
0V
10V
100msec~1sec
100msec~1sec
CLOSE
OPEN
500msec
500msec
100msec~1sec
100msec~1sec
④
Preset operation setting by program number
On the P2 control mode selection screen, change the settings to "RFOP=ANALOG" and
"AMOP=SELECT".
While RF is off, change the program number with the program strobe signal and input the
motor strobe signal to perform a preset operation.
Set the one-shot signal pulse width for the program strobe signal and the motor strobe signal
to a width between 100 ms to 1 s. If the pulse is less than 100 ms or greater than 1 s, stable
operation may not be possible. We recommend using a 500 ms pulse width for the most
stable operation possible.
Figure 5-9 Control mode selection screen (P2)
R F O P = A N A L O G
O S C = I N T
A M O P = S E L E C T
A D L Y =
0 s
Figure 5-10 Timing chart for preset operation by program number