![Ulvac PMB100D Instruction Manual Download Page 34](http://html1.mh-extra.com/html/ulvac/pmb100d/pmb100d_instruction-manual_4492079034.webp)
YK11-0016-DI-002-08
7
※
1) No surface treatment (standard model). Optional surface treatment (Alumite) is available
(PMB***D-T model).
※
2) Measured by Pirani vacuum gauge. It will be about 4.0 x 10-2Pa when measured by MacLeod
vacuum gauge.
※
3) Either 200V class or multiple voltage motor is selectable. Multiple voltage motor is not
available with explosion proof motor.
※
4) ULVOIL
(
R-72
)
is standard oil for air-cooled operation.ULVOIL
(
R-72
)
or fluorine oil (J25F) is
selectable as option.
The type of Vacuum pump oil differs depending on vapor pressure, viscosity, etc.
The use of a wrong type of oil can lower the pump performance.
So use the pump oil recommended ULVAC.Recommended type of oil: ULVOIL R-42 (standard oil),
※
5) Air-cooled operation is possible in all pressure range in the case of inverter equipped type.
Specify the type when air-cooled operation is required. Air-cooled operation is not possible in
the case of the inverter for water-cooled operation. More than 1 hour operation in
300Pa~4000Pa range could be cause of damage in the case of the PMB600D, 1200D and
2400D. Select water-cooled type in that case.
※
6) Make sure no condensation if cooling water temperature is low.
※
7) Weight of D and DT model is same.
※
8) Performance of these pumps could change depending on roughing pump. Above data is
based on standard roughing pump.
※
9) Ultimate pressure is 0.67Pa in case standard roughing pump is GR series.
※
10) It may cause inverter damage by vibration when the pump is directly mounted on the
roughing pump. Select the pump with inverter not equipped
※
11) The value of the horizontal exhaust specification.
※
12) Complied with cTUVus recognition when water cooling operation (except PMB100D) is
selected and optional inverter (atmospheric pressure operation) is not mounted from the
factory.
※
13) Pressure proof explosion-resistant flange motor (d2G4) and increased safety explosion-
resistant motor (eG3) are selectable as option.
※
14) It is not possible to select both explosion proof motor and atmospheric pressure start type
(inverter).
Summary of Contents for PMB100D
Page 2: ......
Page 3: ...Declaration of Conformity...
Page 39: ...YK11 0016 DI 002 08 12 2 4 Dimensional drawing Fig 11 Dimensional drawing PMB100D...
Page 40: ...YK11 0016 DI 002 08 13 Fig 12 Dimensional drawing PMB300D...
Page 41: ...YK11 0016 DI 002 08 14 Fig 13 Dimensional drawing PMB600D...
Page 42: ...YK11 0016 DI 002 08 15 Fig 14 Dimensional drawing PMB1200D...
Page 43: ...YK11 0016 DI 002 08 16 Fig 15 Dimensional drawing PMB2400D...
Page 62: ...YK11 0016 DI 002 08 35...
Page 65: ...YK11 0016 DI 002 08 38...
Page 72: ...YK11 0016 DI 002 08 45...
Page 84: ...YK11 0016 DI 002 08 57 Fig 28 PMB300D Optional attachment figure...
Page 85: ...YK11 0016 DI 002 08 58 Fig 29 PMB600D Optional attachment figure...
Page 86: ...YK11 0016 DI 002 08 59 Fig 30 PMB1200D Optional attachment figure...
Page 89: ...YK11 0016 DI 002 08 62 Fig 33 Dimensional drawing PMB2400D Horizontal exhaust model...
Page 99: ...YK11 0016 DI 002 08 72...
Page 105: ...YK11 0016 DI 002 08 78...
Page 119: ......