To make sure that the service mode does not remain active unintentionally, it is automatically deactivated after
2 hours, and before that, the TriBox3 will begin switching the buzzer on and off once a second to notify the user.
To prevent, for example, unwanted measurements when manually cleaning the sensors and thus potentially
generating false alarms, the TriBox3 box should be set to service mode beforehand.
6.2.2 Checking the Analog Outputs
To check the functioning of the analog outputs, a constant, fixed output current can be set in the menu. This
value can be measured directly at the corresponding analog output (see chapter 5.2). Holding the analog values
is automatically deactivated when the user returns to the main view.
Danger to life due to electric shock! Due to the free mains voltage in the unit,
tests may only be carried out by qualified personnel who are authorised to
do so on the basis of their training. The relevant safety and VDE regulations
must be observed.
DANGER