3. Select “Calibrate” in the menu of the selected sensor.
4. Calibration depends on the type of sensor. After selecting the value to be calibrated, the calibration wizard
starts. The respective reference standards should already be available or prepared for calibration (e.g.
TpH-D reference solutions etc.).
5. The instructions of the wizard must be strictly followed, otherwise the accuracy of the measurements
cannot be guaranteed after completion of the wizard.
6. After calibration, the service mode can be exited.
A room temperature of around 20 °C is required for precise calibration.